Description
Dimensions (W x H x D): 410 × 152 × 368 mm
Weight: 10.0 kg
Electrical
Rated power voltage: 240V/230V/115V/110V/100V± 10%
Rated power frequency: 50/60 Hz
Power current: 4.0 A at 230 – 240 V/8.0 A at 100-115 V
Power consumption in Standby mode: 25 watts
Power consumption at max. HF power: 620 watts 920 VA
Current consumption in Standby mode: 150 mA at 230 – 240V/300 mA at 100-115 V
Potential equalization terminal: Yes
Power fuses: 2 slow burn, 4 A at 230 – 240 V/8 A at 100-115 V
Cutting
HF power adjustment: With up/down increment keys
Form of HF voltage: Sine-shaped
Nominal frequency of HF voltage: 330 kHz
Cutting quality: 4 coagulation effects, selectable by key
Constancy of the coagulation effects: Automatically regulated
Support for initial incision and cutting: PPS, PowerPeakSystem
Monopolar AUTO CUT: With automatic voltage control
Monopolar HIGH CUT: with automatic arc control
Nominal HF output: 300 watts at 500 ohms
Peak HF power with PPS: 400 watts at 500 ohms
Monopolar ENDO CUT: 200 watts at max 200 ohms
BIPOLAR AUTO CUT: With automatic voltage control
Nominal HF output: 100 watts at 500 ohms
Coagulation
HF power adjustment: With up/down increment keys
Monopolar SOFT COAG
Form of HF current: Sine-shaped
Nominal frequency of HF voltage: 300 kHz
Nominal HF output: 120 watts at 125 ohms
End of coagulation: with Pedal or AUTO STOP
Monopolar FORCED COAG
Form of HF current: Pulse-modulated
Nominal frequency of HF voltage: 1 MHz
Nominal HF output: 120 atts at 350 ohms
Monopolar SPRAY COAG
Form of HF current: Pulse-modulated
Nominal frequency of HF voltage: 1 MHz
Nominal HF output: 120 watts at 500 ohms
Biopolar COAG
Form of HF current: Sine-shaped
Nominal frequency of HF voltage: 330 kHz
Nominal HF output: 120 watts at 125 ohms
Start of coagulation: By foot pedal or AUTO START
End of coagulation: By foot pedal or AUTO STOP
Safety Precautions
Safety class: I, corresponds to DIN IEC 601-1
Type: CF, corresponds to DIN IEC 601-1
Software check of LF leakage current, Icritical: 0.05mA (only with capacitive grounding)
Software check of HF leakage current, Icritical: 150 mA/300MA
NESSY: Neutral electrode Safety System
Neutral electrode switching: Switch-selectable capacitively grounded or with floating power
Self-Check: Includes all safety-related subassemblies
Environmental Conditions for Storage
Temperature -40°C to + 70°C
Air humidity, relative 30% to 95%
Environmental Conditions for Operating
Temperature +10°C to + 40°C
Air humidity, relative 30% to 75%, noncondensing